NORMA 6003/NORMA 6003+/NORMA 6004/NORMA 6004+
Users Manual
3. Use the arrow (K L M N) keys to move the cursor to "U
Specify and then push S.
o
If Use Measured is selected, the measured voltage fundamental amplitude is used as the rated
voltage.
o
If Specify is selected, you can specify the rated voltage. Move the cursor to the corresponding
input box, then push S, and input the corresponding value using the on-screen keyboard,
then move the cursor to OK key on the keyboard, and push S.
4. Use the arrow (K L M N) keys and Sto enter the hcf settings.
For I
,
N
o
If Use Measured is selected, the measured current fundamental amplitude is used as the rated
current.
o
If Specify is selected, the rated current can be specified.
5. Use the arrow (K L M N) keys to move the cursor to the Apply button and push S. The
settings are not saved to the device if you do not select Apply.
6. Push J to go back to the previous menu.

Meter

When the Analyzer is turned on, it defaults to the Meter mode, and the current measurements are
displayed in a list on the main screen.
You can also push l to enter the Meter mode.
Table 12
lists the items displayed on the screen.
Item
Range configuration
information
Wiring configuration
information
Update rate
Current state
Value List
30
Table 12. Meter Screen
Current range information for each module
The representation with (Auto) indicates auto range, which
automatically switches the range according to the signal.
The representation without (Auto) indicates that the current range is
fixed, the range of the value displayed and the range will not change
with the signal.
Current wiring selection for each wiring group.
The current update rate of the Analyzer. See
Rate.
Preview: Currently in Preview mode.
User Item: Currently in User Item mode.
HOLD: Currently in data hold state. The value displayed in the value
list area does not change with the actual signal.
The measurement items are displayed in a list according to the user's
settings. See
Preview
" (rated voltage) select Use Measured or
N
Description
and
User Item View
for details.
Set Sync Source and Update