Output Option Card - Badger Meter Data Industrial 3050 Series Installation & Operation Manual

Btu monitor
Hide thumbs Also See for Data Industrial 3050 Series:
Table of Contents

Advertisement

Series 3050 Btu Monitor

OUTPUT OPTION CARD

If the Series 3050 Btu Monitor was ordered with the Output Option card, it will have several additional outputs.
1. Analog Output ( 0-20mA; or 4-20mA ) which can be converted externally to 0-5VDC, 1-5VDC with a 250 Ohm resistor; or
0-10VDC or 2-10VDC with a500 Ohm resistor. A 15VDC power supply is provided to permit current sinking or sourcing. The
Series 3050 has special software that permits the Analog output.
2. USB for direct access to a computer using a standard mini-USB cable.
3. RS-485 for fully addressable Modbus or BACnet communication.
Page 14
1 RS485 B
2 RS485 A
3 RS485 GND
4 LOOP +
5 LOOP -
6 GND
Figure 12: Current Sourcing Analog Output
1 RS485 B
2 RS485 A
3 RS485 GND
4 LOOP +
5 LOOP -
6 GND
Figure 13: Current Sinking Analog Output
Figure 12
RS485 Communication
1 RS485 B
2 RS485 A
3 RS485 GND
4 LOOP +
5 LOOP -
6 GND
Figure 14: RS485 Communication (Modbus and BACnet)
Figure 13
Model 3700
Model 345WT
or other
MODBUS
Master
Device
Shield
RS485 -
RS485 +
6-11