Connecting A 600-Series Pump - Waters 2475 Operator's Manual

Multi fluorescence detector
Table of Contents

Advertisement

Connecting a chart recorder to the detector
+
Chart recorder
terminals
Chart marks
You can also generate a chart mark from the detector's front panel whenever
you perform these actions:
Press Chart Mark on the detector's keypad.
You program a timed event to generate a chart mark.
A chart mark signal is generated at the chart mark inputs on the analog
connector.

Connecting a 600-series pump

To connect a 600-series pump:
1.
Place the detector on a level surface.
2.
Make the plumbing connections described on
2475 detector
B (inputs and outputs)
Red
Black
+
1
Detector out 1
2
Detector out 1
3
Ground
+
4
Detector out 2
5
Detector out 2
+
6
Switch 1
7
Switch 1
8
Ground
+
9
Switch 2
10
Switch 2
page
2-3.
Connecting other devices
2-29