Instrument Testing
When the test is completed, the results report appears. Scroll through the
8.
report; every result should show PASS. If any result shows "FAIL", try the
following and rerun the test. If the test continues to fail, contact Technical
Support.
 Make sure the information entered into Gen5 matches the Test Plate's
Certificate.
 Verify that the Test Plate is within its calibration certification period. If it is
out of date, contact Technical Support to schedule a recertification.
 Ensure that the Test Plate is correctly seated in the microplate carrier.
 Check the alignment (corner) holes on the Test Plate to ensure they are clear
of debris.
 Check the filters on the Test Plate to ensure they are clean. If necessary,
clean them with lens paper. Do not remove the filters from the test plate. Do
not use alcohol or other cleaning agents.