Establish Communication; Troubleshooting; Verify Performance - Agilent Technologies BioTek Epoch 2 Instructions For Use Manual

Absorbance microplate spectrophotometer
Table of Contents

Advertisement

Installation

Establish Communication

On the host computer, start Gen5 and log in if prompted.
1. From the main screen, select
2. Set the Reader Type to
3. Perform one of the following steps, as applicable:
• Select
• Set the
connected.
4. To verify that Gen5 can communicate with the instrument, click

Troubleshooting

If the communication attempt is successful, Gen5 displays a success message. Return to
the main screen. If the communication attempt is not successful, try the following:
• Is the reader connected to the power supply and turned on?
• Is the communication cable firmly attached to both the reader and the
computer?
• Did you select the correct Reader Type/instrument in Gen5?
• Try a different COM Port or use Plug & Play.
• Did you install the USB driver software?
If you remain unable to get Gen5 and the reader to communicate with each other,
contact Technical Support.

Verify Performance

Refer to the Instrument Testing section for Absorbance Test procedures.
Agilent BioTek Epoch 2 Instructions for Use
1771011I Revision C
System > Instrument Configuration
Epoch
2.
Plug &
Play.
Com Port
to the computer's COM port to which the reader is
Page 20 of 51
and click Add.
Test
Comm.
Agilent Technologies, Inc.

Advertisement

Table of Contents
loading

Table of Contents