Thermo Scientific
'P' COMMAND
Repeat: YES
Frequency: Every two seconds
Compile Dependencies: NONE
Description: This command displays the Ports of the 8051 microcon-
troller in the following fashion: "P0-XX P1-XX P2-XX P3-XX" (the XX
are hexadecimal values)
'R' COMMAND
Repeat: NO
Frequency: N/A
Compile Dependencies: NONE
Description: This command displays the contents of the 8051 internal
RAM in the following fashion: "RAM (0-255 decimal) (0x00-0xFF
hexadecimal) (value in hexadecimal)".
'S' COMMAND
Repeat: NO
Frequency: N/A
Compile Dependencies: NONE
Description: This command reads the serial number (if the unit does not
have one, it reads the locations where the serial number would be stored)
in the following fashion: "SN Ext RAM (0-15 decimal) (0x00-0x0F
hexadecimal) (value in hexadecimal)".
't' COMMAND
Repeat: YES
Frequency: Every two seconds
Compile Dependencies: Env Chamber, Incubator
Description: This command display miscellaneous temp information.
INCUBATORS:
Param 1: Temperature setpoint (divide by 10 to convert to °C)
Param 2: Temperature warm alarm setpoint (divide by 10 to convert to
°C)
Param 3: Temperature cold alarm setpoint (divide by 10 to convert to
°C)
Param 4: Cabinet Probe RTD including TempOffset (divide by 10 to
convert to °C)
Param 5: Water RTD (divide by 10 to convert to °C)
Param 6: Door integral
Param 7: Water integral
Param 8: High temperature excursion (divide by 10 to convert to °C)
Param 9: Low temperature excursion (divide by 10 to convert to °C)