IntrLogic Power/Key
On Modes
Thermo Scientific

Appendices

Section 8
All power/key on modes require that the keypad pattern be selected by the
time the logo stops scrolling, and that the pattern be maintained for the
approximate three seconds in which the logo is stationary in the center of
the display until the logo disappears. It should also be noted that on a first
time power up sequence for a CPU board (defined as the Battery Check
Interval variable upon restoration from the non-volatile memory being
something other that 4, 12, or 20), all keypad patterns are ignored and an
automatic start-up sequence ensues. In addition, normal start-up will
occur with no keypad pattern.
Abbreviations:
Decrement Pad - Dec
Increment Pad - Inc
Mode/Mute Pad - Mode
Table 7-1. Keypad Patterns
Keypad Pattern
Description
Incubator
Dec
Adjust the Water Adj (ADJ Delta H2O variable)
Inc
Plug-N-Play mode
Mode
N/A
Override serial number requirement or permit serial number reprogram
Mode + Inc
(code must be compiled with SN Init And Test)
Mode + Dec
N/A
Inc + Dec
Restore Defaults / Manual Reset
Mode + Inc + Dec
Board Test Mode (if compiled with UL Testing, then UL Test Mode)