Midi 40 CO2 Incubator
'M' COMMAND
Repeat: N/A
Frequency: N/A
Compile Dependencies: NONE
Description: The defined way to stop telemetry output. It will not stop
error messages, just the telemetry table data. To guarantee that no
telemetry messages are sent, set JP3-TELE jumper on the CPU board to
RS485 mode (short pins 2 & 3).
'O' COMMAND
Repeat: YES
Frequency: Every two seconds.
Compile Dependencies: Available in all code - effected by Incubator and
O2 Control
Description: This command lists the offsets.
Param 1: Temperature offset (divide by 10 to convert to °C)
Param 2: Relative humidity offset (divide by 10 to convert to %)
Param 3: Incubator only...CO2 offset (divide by 10 to convert to %)
Param 4: O2 Control only...O2 offset (divide by 10 to convert to %)
'p' COMMAND
Repeat: YES
Frequency: Every two seconds.
Compile Dependencies: Env Chamber, Incubator, CO2 Control, O2
Control, RH Control
Description: This command displays PWMs and timers for various con-
trol sets.
INCUBATORS:
Param 1: Door Heater PWM
Param 2: Water Heater PWM
Param 3: CO2 injection PWM
Param 4: CO2 rapid recovery time (used only in rapid recovery mode)
Param 5: O2 Control only - O2 injection PWM
Param 6: O2 Control only - O2 rapid recovery time (used only in
rapid recovery mode)
Param 7: RH Control only - Relative humidity heater PWM
Param 8: RH Control only - Relative humidity water source PWM