AGFA CR 30-X User Manual page 68

Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

CR 30-X
Condition
Digitizer cannot
take the image
plate with the
transport unit
(e.g. cassette fell
off, IP was not
inserted
correctly)
Communication
error, FireWire is
not plugged
Time-out after 5
minutes when no
ID data was
entered in Fast ID
mode
Messages will stay active until the problem is solved or the dialog box message
on the control PC is confirmed by clicking the OK button.
42
Operating CR 30-X
Control PC message
"Digitizer cannot take the
image plate with the
transport unit. The cassette
may be damaged. Use
another cassette. Call your
local service!"
"Communication-error.
Check if fire wire
connection between
digitizer and Control-PC."
"No ID data provided
within the timeout period.
Image plate was returned
without scanning."
Status
indicator
Red, flashing
After
Use another cassette
unlocking:
and call Agfa service.
green,
constant
Check if FireWire is
Red, flashing
plugged in or call Agfa
service.
Confirm the message
on the PC and remove
After
the cassette.
unlocking:
green,
To scan this cassette
constant
insert it in the digitizer
again.
Action