CR 30-X P
C
LATES AND
ASSETTES
Put the black tray with the image plate onto the cassette, as shown in the figure
5
below.
The tray has 2 little hooks. Put the tray onto the cassette in such a way that the hooks
reach over the border of the cassette. This to avoid a bending of the tray and image
plate.
Clean the image plate and the tray as follows:
6
Moisten the cloth with cleaner and wipe the image plate surface softly
and evenly (phosphorous side and rear side).
Leave the image plate surface to dry for about 10 minutes to allow the
solvents to evaporate.
Use only a dry cloth or blow away possible dust with compressed air
(canned air), to clean the tray thoroughly. (DO NOT USE A SCREEN
CLEANER TO CLEAN THE TRAY.)
Once the image plate surface is dry (after about 10 minutes), check once
again for particles of material and other impurities before placing the tray
and image plate set into the cassette.