Specifications
Table 3–8: Environmental Characteristics
Characteristics
Temperature
Nonoperating
Operating
Altitude
Nonoperating
Operating
Vibration (Operating)
Shock
Transportation
Table 3–9: Certifications and compliances
Category
EC Declaration of Conformity –
1
EMC
FCC Compliance
Installation (Overvoltage)
Category
3–18
Performance Requirements
–40 to +65 C (–40 to +149 F)
0 to +50 C (32 to 122 F)
To 50,000 ft (15,240 m)
To 15,000 ft (4,572 m)
Fifteen minutes each axis at 0.025 inch.
Frequency varied from 105510 Hz in
4-minute cycles with the instrument
secured to the vibration platform; ten
minutes each axis at any resonant point or
at 55 Hz.
1
50 G's,
/
sine, 11 ms duration, three
2
guillotine shocks per side.
Qualified under NTSA Test Procedure 1A,
Category II (36-inch drop).
Standards or description
Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility. Compliance was
demonstrated to the following specifications as listed in the Official Journal of the European Union:
EN 50081-1 Emissions:
EN 55022
EN 50082-1 Immunity:
IEC 801-2
IEC 801-3
IEC 801-4
1
High-quality shielded cables must be used to ensure compliance to the above listed
standards.
Emissions comply with FCC Code of Federal Regulations 47, Part 15, Subpart B, Class A Limits.
Terminals on this product may have different installation (overvoltage) category designations. The
installation categories are:
CAT III
Distribution-level mains (usually permanently connected). Equipment at this level is
typically in a fixed industrial location.
CAT II
Local-level mains (wall sockets). Equipment at this level includes appliances, portable
tools, and similar products. Equipment is usually cord-connected.
CAT I
Secondary (signal level) or battery operated circuits of electronic equipment.
Class B Radiated and Conducted Emissions
Electrostatic Discharge Immunity
RF Electromagnetic Field Immunity
Electrical Fast Transient/Burst Immunity
Supplemental Information