9-3 Other Search Algorithms
9-3-1 Variation Search
This algorithm searches the points where the voltage reaches its minimum and maximum
values. It sets these points as origins to detect the points where the variations begin to be
continuously within the specified range respectively in minimum and maximum voltage
directions. It then determines the period between these detected points as the passage time.
①
Time limit for searching a maximum voltage point from the minimum voltage point: Up to 200ms
②
Duration for which variations are not searched in the minimum voltage direction
③
Duration for which variations are not searched in the maximum voltage direction
④
Duration for which the variation condition continues to be satisfied in the minimum voltage direction
⑤
Duration for which the variation condition continues to be satisfied in the maximum voltage direction
⑥
Variation condition to be continuously satisfied in the minimum voltage direction (mV or less)
⑦
Variation condition to be continuously satisfied in the maximum voltage direction (mV or less)
4500
4000
3500
3000
2500
2000
1500
Search invalid duration
in the minimum voltage direction
1000
500
0
Passage start point
④
Variation condition satisfied duration
⑥
Contiguous condition
satisfied variations
②
0
0.05
①
②
④
⑥
Passage end point
⑤
Variation condition satisfied duration
Contiguous condition satisfied variations
⑦
③
Search invalid duration
in the maximum voltage direction
Time elapsed between the minimum voltage
point and the maximum voltage point
0.1
経過時間 [ms]
Elapsed time [ms]
36
③
⑤
⑦
①
0.15
0.2