Mounting
2.3

Rotate the electronics on the meter (optional)

You can rotate the transmitter on the meter up to 90°.
1.
2.
3.
2.4
Rotate the display on the transmitter
(optional)
The display on the transmitter electronics module can be rotated 90° or 180° from the
original position.
20
Using a 4 mm hex key, loosen the cap screw that holds the transmitter in place.
Figure 2-4:
Component to secure transmitter in place
A. M5 socket-head cap screw
Rotate the transmitter clockwise to the desired orientation up to 90°.
Secure the cap screw in place and tighten to 60 lb·in (6.8 N·m).
A
Micro Motion
®
Gas Density Meters