Technical Specifications - Presys TA-60NL Technical Manual

Temperature advanced calibrators
Table of Contents

Advertisement

PRESYS
 Instruments
Another way of performing automatic documented calibrations is by means of
Calibration Software for PC/Windows
connect the computer.
TA Calibrators have also many other features, such as:
 RTD input for 2, 3 and 4 wires. Table IEC 60751, JIS or Callendar-Van Dusen
user-configurable. Engineering units configurable to °C, °F and K.
 Present inputs for mA, thermocouples and thermo-switches.
 Thermo-element reading scaled to ITS-90 or IPTS-68.
 Use of internal standard thermometer.
 Accuracy to 0.1 ºC, stability of 0.025 ºC and resolution of 0.01ºC.
 Carry out completely automatic calibrations without the use of a computer.
 Built in Web Server, Ethernet communication and USB serial Communication.
 USB port for software/firmware upgrade.
 HART
®
Communication Protocol (optional) with internal resistance configurable,
transmitter power supply and latest DD as option.
 The electric signal calibrator is independent from the dry block function.
 Display indication when the temperature reaches the desired value.
 5.7 inches touchscreen display that eases the operation and configuration of the
calibrator.
 Internal regulated 24 Vdc power supply for 2-wire transmitters.
 Independent circuit for over-temperature protection and safety.
 Insert to choose, handles and test leads included.

1.1 - Technical Specifications

Operating Range
Power Supply
Well Diameter / Depth
Resolution
Display Accuracy
Stability (after 10 min)
Axial Uniformity
(40 mm)
Radial Uniformity
Power Consumption
Heating Rate
Cooling Rate
Dimensions (H,W,D)
Weight
(1) Ambient Temperature up to 23 ºC.
Note: The times for cooling and heating rate presented refer exclusively to the use of the
solid (metallic) insert. For use with stirred liquid insert the time may vary according to the
fluid used and its viscosity.
Page 2
TM
, which uses USB communication or Ethernet to
-60 ºC to 140 ºC
115 Vac or 230 Vac 50/60 Hz, according to order code.
 35 mm x 160 mm
0.01 ºC (0.01 ºF)
 0.04 ºC (full range)
 0.02 ºC (full range)
30 minutes (25 ºC to 140 ºC)
60 minutes (25 ºC to -60 ºC)
370 x 306 x 450 mm
TA-60NL
(1)
0.10 °C
 0.025 ºC
870 W (rated)
17 kg
TA-60NL