Note: To enable the Scale Function, turn the function ON before touching
OK button. To disable, turn the function OFF.

2.1.4 - Saving Current Configuration (Memory Manager)

The TA Series calibrators admit several special functions that may become of
frequent use. In these situations, it is useful to store such settings in the instrument in
order to save time.
After setting the desired calibration mode (input type, probe reference or special
function), press the icon
option CREATE NEW can be given a name for this configuration and a description.
Press the SAVE button.
The operation that was being performed by the TA Calibrator shall be stored in
memory identified by the name given to it. To use it again, even after the calibrator is
turned off and on, select the name of the desired setting and press the LOAD button.
The SAVE AS DEFAULT button sets the current configuration as the default
configuration of the calibrator. Thus, every time the calibrator is turned on, this will be the
initial configuration of the calibrator.
Page 20
Fig. 18 - Scale Function Configuration
> MENU, and the button MEMORY MANAGER. On the