2.1.2 - Input Settings

The INPUT menu has the following options:
For OHM measurement, you should also select between 2, 3 or 4 wires options.
For RTD input, it should be chosen the type between Pt-100, Pt-1000, Cu-10 or
Ni-100 (standard table used), the number of lead wires (2, 3 or 4 wires) and the
temperature scale (ITS-90 or IPTS-68). There is also the option to configure the
Callendar-Van Dusen coefficients of the sensor, selecting the option CVD and the
desired curve in the list.
To create new CVD coefficients, press
appear in the created list as identified in ID.
For TC (thermocouple), you must select the thermocouple type and the type of
cold junction compensation (CJC): Internal or Manual. In Internal option, the
Page 16
Fig. 11 - Input Menu Options
Fig. 12 - Options for RTD Input
(edit), and ADD button. The curves