2.3 - Automatic Tasks

In TA Calibrators, automatic calibration tasks can be created and performed. This
option can be used to generate calibration work orders for sensors, transmitters and
indicators.

2.3.1 - Creating Tasks

To create tasks from the main menu, select CALIBRATOR. Select the desired
type of input and probe configuration. For example, to calibrate a temperature
transmitter, select the probe configuration (Internal or External Reference) and mA input
(which will be connected to the current output of the transmitter). For a temperature
indicator, e.g., selected NONE for the input.
Press the
Fill at least the serial number of the instrument to calibrate, instrument TAG, lead
time (time, in seconds, for the complete stabilization of the system), maximum error
allowed for the instrument to be calibrated (in % of the span, reading or full scale),
calibration range (highlighted fields). For the SWITCH tasks, the error fields are filled in
the next step.
icon, and select TASKS and CREATE NEW TASK.