The TA Calibrator automatically starts the execution of the calibration, generating
the registered setpoints, waiting the stabilization of the system and reading the
associated input values. If you selected the option NONE as input, for each generated
point, the calibrator will require to digit the value of the temperature indicated by the
instrument (thermometer, indicator...). The result will be displayed on the screen, and a
progress bar indicates the calibration remaining time. At the end of the calibration, a
report is showed with the generated values, the obtained values, the expected values,
and the error. If the error is higher than the registered value for the task, the line appears
in red.
The first time that a task is performed, it will be saved as As-found (before
adjustment). If it runs again, it will be saved as As-left (after adjustment). The results are
saved in the calibrator and can be viewed at any time.

2.3.3 - Viewing Results

Once a task has been performed, it remains saved in the calibrator.
To view the results of a calibration by the calibrator, in the main menu select
TASKS.
Enable the option ● PERFORMED. The list will show only the tasks that have
been performed. Select the desired work order and press OK. On screen, the report with
the calibration points, the obtained values, expected values and the errors will be shown.
If the error is higher than the value registered for the task, the line appears in red.
The task data is saved in a PDF file in the internal memory card of the Calibrator
and can be accessed connecting the calibrator to a computer. To save the data in a
PenDrive or External HD on the host USB, press the PenDrive icon
Page 28
Fig. 30 - Task Results