Obtaining The Test Point Values; Evaluating The Test Points - Emerson Micro Motion 2400S Configuration And Use Manual

Transmitter for devicenet
Hide thumbs Also See for Micro Motion 2400S:
Table of Contents

Advertisement

Troubleshooting
11.19.1

Obtaining the test point values

To obtain the test point values:
With the display, configure the required test points as display variables. See Section 8.9.5.
With ProLink II
a. Click
b. Observe or record the values displayed for
and
Drive Gain
With a DeviceNet tool, execute a Get for the attributes listed in Table 11-4.
Table 11-4
Test points with DeviceNet tool
Test point
Drive gain
Tube period
Left pickoff
Right pickoff
11.19.2

Evaluating the test points

Use the following guidelines to evaluate the test points:
If the drive gain is erratic, negative, or saturated, refer to Section 11.19.3.
If the value for the left or right pickoff does not equal the appropriate value from Table 11-5,
based on the sensor flow tube frequency, refer to Section 11.19.4.
If the values for the left and right pickoffs equal the appropriate values from Table 11-5, based
on the sensor flow tube frequency, record your troubleshooting data and contact Micro Motion
customer service. See Section 11.3.
Table 11-5
Sensor pickoff values
(1)
Sensor
®
ELITE
CMF sensors
F025, F050, F100 sensors
F200 sensors
H025, H050, H100 sensors
H200 sensors
R025, R050, or R100 sensors
R200 sensors
T-Series sensors
CMF400 I.S. sensors
(1) If your sensor is not listed, contact Micro Motion. See Section 11.3.
Configuration and Use Manual
ProLink
>
Diagnostic Information
.
Class
Diagnostics Object (0x66)
Pickoff value
3.4 mV peak-to-peak per Hz based on sensor flow tube frequency
3.4 mV peak-to-peak per Hz based on sensor flow tube frequency
2.0 mV peak-to-peak per Hz based on sensor flow tube frequency
3.4 mV peak-to-peak per Hz based on sensor flow tube frequency
2.0 mV peak-to-peak per Hz based on sensor flow tube frequency
3.4 mV peak-to-peak per Hz based on sensor flow tube frequency
2.0 mV peak-to-peak per Hz based on sensor flow tube frequency
0.5 mV peak-to-peak per Hz based on sensor flow tube frequency
2.7 mV peak-to-peak per Hz based on sensor flow tube frequency
.
Tube Frequency
,
Left Pickoff
Instance
1
,
Right Pickoff
,
Attribute
20
21
23
24
123