2.2. Securing the Stage to the Mounting Surface

DANGER: Mechanical Hazard!
Personnel must be made aware of the mechanical hazards during set up or when you
do service to the stage.
Do not manually move the stage if it is connected to a power source.
l
The stage must be mounted securely. Improper mounting can result in injury and
l
damage to the equipment.
When you move the tabletop manually to do maintenance, this could expose the
l
operator to pinch points. Refer to
DANGER: Magnetic Field Hazard! When you move the tabletop manually to do
maintenance, this could expose the operator to the magnet track. Refer to
Hazards
The mounting surface must be flat and have adequate stiffness to achieve the maximum
performance from the stage. When it is mounted to a non-flat surface, the stage can be distorted
while the mounting screws are tightened. This distortion will decrease overall accuracy. Adjustments
to the mounting surface must be done before the stage is secured.
Inspect the mounting surface for dirt or unwanted residue and clean if necessary. Use precision
flatstones on the mounting surface to remove any burrs or high spots. Clean the mounting surface
with a lint-free cloth and acetone or isopropyl alcohol and allow the cleaning solvent to completely
dry. Gently place the stage on the mounting surface.
IMPORTANT: The stage is precision machined and verified for flatness prior to product
assembly at the factory. If machining is required to achieve the desired flatness, it
should be performed on the mounting surface rather than the stage. Shimming should
be avoided if possible. If shimming is required, it should be minimized to retain
maximum rigidity of the system.
Table 2-1:
Stage Mounting Surface Flatness Requirement
Stage Travel
All Travels
1. Verify that the shipping clamps are secure.
2. Remove the pneumatics cover. It is attached to the vertical angle bracket with four M3 button
head screws.
3. Mount the Z-Axis assembly using one of the available customer mounting patterns shown in
Section 2.1.
You can manually, but gently, move the base of the pneumatics cylinder to the left or
right to gain access to mounting holes in the vertical angle bracket. Use a long ball end T-handle.
a. NOTE: Do not use the two mounting holes located directly below the pneumatics cylinder
b. Refer to
Table 2-2
4. Reattach the pneumatics cover by reversing steps used to remove it.
WARNING: Mechanical Hazard! Do not operate the stage without the pneumatic
cover in place.
22
on
Page
13.
for screw torque specifications.
Pinch Points
on
Page
12.
Hardware Manual
Magnetic
Flatness Requirement
7.6 µm
www.aerotech.com