are stored in RAW files. The actual calibration process is done after the data collection
is completed.
The basic calibration of scale and yaw is made by comparing the Course Over Ground
(COG) and Speed Over Ground (SOG) data from the current indicator and the GPS.
Note
Velocity measurement calibration is always done using previously recorded raw data.
You must collect data from four survey lines. The data from each survey line can be
provided in any number of
The information provided by the calibration is entered into the echo sounder software as
operational parameters. This is done automatically.
Calibration Wizard
The
Calibration Wizard
process.
1
The first page in the
process, or return to a previous process by reusing saved data.
2
The second page in the
calibrated.
3
The third page in the
into the calibration process.
Related topics
Calibrating the current velocity measurements, page 28
Functions and dialog boxes, page 35

Calibration Wizard dialog box #1

The first page in the
process, or return to a previous process by reusing saved data.
Prerequisites
The
Calibration Wizard
How to open
To open the
Calibration Wizard
Replay mode, then select
36
files.
*.raw
offers a set of dialog boxes to guide you through the calibration
Calibration Wizard
Calibration Wizard
Calibration Wizard
allows you to either start a new calibration
Calibration Wizard
can only be opened from within the calibration process.
, place the echo sounder system in either Normal or
on the
Calibration
allows you to either start a new calibration
allows you to select the channel to be
allows you to import the calibration data
menu.
Setup
110-0002398/A