NOTE: Shaker must be in TIME mode — TIME signal must be on.
6.0 Troubleshooting
Problem
POWER switch does not light
Shaker stalls – Message Er1
Display does not light
Message Er2, Er3, or Er4 on display
Should you have a question about the operation of the Labnet Orbit™ Digital Shakers or if service is required, contact your Corning
representative. Do not send in a unit for service without first calling to obtain a repair authorization number. Should the unit require
return to Corning for service, it should be properly packed to avoid damage. Any damage resulting from improper packaging shall be
the responsibility of the user.
7.0 Maintenance and Cleaning
Before cleaning the device, unplug the main cord from the wall socket.
With the exception of the occasional maintenance described in Section 4.0 of this manual, no scheduled maintenance is normally
required. However, an experienced technician should regularly check the device operation at least once a year to make sure it is
operating correctly.
Regular cleaning of the housing is recommended.
The housing of the shaker can be cleaned with special cleaners for polyurethane (plastic) surfaces. A damp (not wet) cloth is
recommended.
NOTE: Do not use any aggressive or abrasive cleaners (acetone, nitro, polish, etc.) because its surface can be permanently damaged.
POWER
RUN
TIME
99.5
Solution
Check the power source.
Check fuses.
Power the unit down, then restart.
Contact Corning Customer Service.
RPM
Encoder
Push
START
STOP
Instruction Manual
|
7
Need help?
Do you have a question about the Orbit S2020-P4-B and is the answer not in the manual?
Questions and answers