13.1.2 OUT commands

OUT commands set device parameters. Remote control mode must be active.
Parameter
Parameter
settings
out_sp_00
xxx.xx
out_sp_03
xxx.xx
out_sp_04
xxx.xx
out_sp_06
xxx.xx
out_sp_07*
x
out_sp_10
xxx.xx
out_sp_11
x
out_sp_27
xxx
out_sp_28
xxx.xx
out_sp_29
xxx.xx
* Do not use for reprogramming, use out_sp_27 instead. Pump stage 4 has an output pressure of
approx. 450 mbar.
Device modes
Parameter
out_mode_01
x
out_mode_03
x
out_mode_04
x
out_mode_05
x
www.calcert.com
15
10
20
5
25
0
30
Setting
Setting for the setpoint temperature
Setting for the high temperature
Setting for the low temperature
Setpoint temperature setting via watchdog function
Set pump to preset levels 1...5
Setting for the actuating variable via the serial
interface
Setting the temperature units:
0 = °C
1 = °F
Pump capacity in %
High temperature alarm limit setting
Low temperature alarm limit setting
Setting
Setting 0 (for reasons of downwards compatibility
the parameters 1 or 2 can also be used. Within the
device it has no effect)
Setting of the operating mode EPROG:
0 = 0-10 V
1 = 0-20 mA
Setting of the temperature control:
0 = Internal control
1 = External control
Start/stop command of the device in remote control
mode:
0 = Stop tempering
1 = Start tempering