8.6.3 Adjust controller
In the menu <Adjust controller> the individual control parameters for
internal and external control can be set.
The device is switched on.
1. Bring up the <Main menu>.
2. From the <Thermodynamics> submenu, select the menu option
<Adjust controller>.
The dialog box for setting the controller appears on the display.
3. Activate internal or external control.
This results in the respective adjustable control parameters. External control
is only possible with an externally connected Pt100 temperature sensor.
4. Switch to the submenu <Control performance>.
5. In the dialog box, set the desired parameters and confirm your entry.
6. For internal control: Activate the aperiodic temperature sequence or the
standard temperature sequence.
The control parameters are set and immediately active.
8.6.4 Limit cooling capacity and heating capacity
In the menu <Limits> the cooling capacity and heating capacity are limited.
The device is switched on.
1. Bring up the <Main menu>.
2. From the <Thermodynamics> submenu, select the menu option
<Adjust controller>.
The dialog box for setting the controller appears on the display.
3. Switch to the submenu <Limits>.
4. In the dialog window that appears, set the limits for cooling capacity and
heating capacity and confirm your entry.
The values are set and are immediately active.
8.6.5 Adjust pump
In the submenu <Adjust pump> the working mode is set, as is the pump
power in one percent increments.
The unit is switched on.
1. Bring up the <Main menu>.
2. From the <Thermodynamics> submenu, select the menu option
<Adjust pump>.
3. Enter the desired value for pump capacity in percent and confirm your input.