Operation

8.6.3 Adjust controller

In the menu <Adjust controller> the individual control parameters for
internal and external control can be set.
 The device is switched on.
1. Bring up the <Main menu>.
2. From the <Thermodynamics> submenu, select the menu option
<Adjust controller>.
 The dialog box for setting the controller appears on the display.
3. Activate internal or external control.
 This results in the respective adjustable control parameters. External control
is only possible with an externally connected Pt100 temperature sensor.
4. Switch to the submenu <Control performance>.
5. In the dialog box, set the desired parameters and confirm your entry.
6. For internal control: Activate the aperiodic temperature sequence or the
standard temperature sequence.
 The control parameters are set and immediately active.

8.6.4 Limit cooling capacity and heating capacity

In the menu <Limits> the cooling capacity and heating capacity are limited.
 The device is switched on.
1. Bring up the <Main menu>.
2. From the <Thermodynamics> submenu, select the menu option
<Adjust controller>.
 The dialog box for setting the controller appears on the display.
3. Switch to the submenu <Limits>.
4. In the dialog window that appears, set the limits for cooling capacity and
heating capacity and confirm your entry.
 The values are set and are immediately active.

8.6.5 Adjust pump

In the submenu <Adjust pump> the working mode is set, as is the pump
power in one percent increments.
 The unit is switched on.
1. Bring up the <Main menu>.
2. From the <Thermodynamics> submenu, select the menu option
<Adjust pump>.
3. Enter the desired value for pump capacity in percent and confirm your input.