Appendix

13 Appendix

13.1 Interface commands

Interface commands allow the device to be remote controlled. Parameters can
be retrieved and the current status can be queried. To do this, the device must
be connected to the master computer via a digital interface. Interface
commands are entered using a terminal program.
Interface commands are divided into IN commands and OUT commands.
IN commands: Retrieve parameters
Command structure: Command + 
E.g. Retrieve the setpoint temperature:
IN_SP_00
E.g. Response of the device:
55.5  LF
OUT commands: Set parameters (only in remote control mode)
Command structure: Command +  + Parameter 
E.g. Set the setpoint temperature to 55.5 °C:
OUT_SP_0055.5

13.1.1 IN commands

IN commands retrieve device parameters.
Process values
System response
in_pv_00
Actual value
in_pv_01
Current variable (%)
In_pv_02
Current temperature of the external Pt100 temperature sensor
in_pv_03
Current temperature of the temperature safety sensor
in_pv_04
Current setting of the high temperature safety function
in_pv_06
Current pressure. EPROG input must be set to pressure
in_pv_07
Current flow. EPROG input must be set to flow
in_pv_16
Current bath fluid filling level (%)