Activate Autostart Function; Set Physical Units; Set Key Tones - Julabo DYNEO DD Operating Manual

Circulator
Table of Contents

Advertisement

1.888.610.7664

8.4.3 Activate autostart function

The autostart function makes it possible to start a temperature control
application directly using the mains switch or via an intermediate timer.
The device is configured ex works in such a ways that it switches to a safe
operating status in the event of power failure. The autostart function is
deactivated. The display shows "OFF." The refrigeration aggregate, heater, and
pump motor are disconnected from the mains voltage.
 The device is switched on.
1. Bring up the <Main menu>.
2. From the <Settings> submenu, select the menu option <Autostart>.
3. Check the <On> box and confirm your setting.
 The autostart function is activated. The next time the device is switched on,
the temperature control application will immediately start with the preset
values. You can also connect and program a timer. In this case the mains
switch must remain on.

8.4.4 Set physical units

The physical units for temperature, pressure and flow can be set in the menu.
 The device is switched on.
1. Bring up the <Main menu>.
2. From the <Settings> submenu, select the menu option <Units>.
3. Select the check boxes to display the desired units for temperature, pressure
and flow.
 The physical units are set.

8.4.5 Set key tones

The device key tones can be enabled and disabled in the menu.
 The device is switched on.
1. Bring up the <Main menu>.
2. From the <Settings> submenu, select the menu option <Key tones>.
3. Enable or disable the key tones.
 The key tones are set. The setting is applied directly.
www.calcert.com
15
10
20
5
25
0
30
Operation
59
sales@calcert.com

Advertisement

Table of Contents
loading

Table of Contents