8. In the <Final state> field, select how the device should behave after the
profile series expires.
9. Enable the <Active> field to activate the profile series.
The profile series is programmed and active. It starts at the programmed
start time on the specified start date and runs through the specified number
of temperature control applications on the specified weekdays. At the end
of series production, the device switches to the defined final state.
8.9
Configure signal outputs
8.9.1 Configure signal outputs of the REG/EPROG plug
The REG/EPROG socket provides three channels as signal outputs. The voltage
or current signals emitted via these channels can be externally processed.
The following values can be output:
•
Setpoint: Specified setpoint
•
Internal: Internal temperature sensor value
•
External: External Pt100 temperature sensor value
•
Performance: Value of the variable
The device is switched on.
The slide-in module with analog connections is installed.
1. Bring up the <Main menu>.
2. Bring up the <Connect unit> submenu.
3. From the <Analog interfaces>/<Analog module> submenu, select the
menu option <Channel 1...3>.
4. In the drop-down window, select the output variable Setpoint, Internal,
External or Capacity.
5. Enter the lowest output value as the 0 V value.
6. Enter the highest output value as the 10 V value.
7. Repeat steps 5 and 6 for channel 2 and channel 3 respectively.
Channel 3 configures the current output. 0 mA or 4 mA can be selected for
the lowest output value.
The signal outputs of the REG/EPROG socket are configured.