Servicing and care
8
42
Fault
Possible cause
Error message:
11 month since last calibration
Calibration is
due
Error message:
Channel temperature > 65 °C or
Device over-
Battery temperature > 60 °C or
heated
any other temperature > 75 °C
Error message:
Abs. Channel pressure
Channel pres-
> 2.05 bar
sure too high
Error message:
Abs. ambient pressure
Ambient pres-
> 1250 mbar
sure too high
Error message:
Abs. ambient pressure
Ambient pres-
< 490 mbar
sure too low
Error message:
Ext DP sensor < -250 mbar or
External differen-
Ext DP sensor > 250 mbar
tial pressure too
high
The display
Flow channel dirty
•
values of the
Calibration due
•
FlowMeter do
Error in setup measurement
•
not match the
device under
test (DUT)
Touch screen
Software locked up
not responsive
Trouble shooting
Contact IMT Analytics soon for calibra-
tion:
www.easy-cal.com
Move device to cooler environment to
avoid damage.
Any further increase in temperature will
damage the device!
Reduce channel pressure below 2 bar
absolute pressure to avoid damage.
Any further increase in pressure will
damage the device!
Reduce ambient pressure below
1200 mbar absolute pressure to avoid
damage.
Any further increase in pressure will
damage the device!
Increase the ambient pressure above
540 mbar absolute pressure to avoid
damage.
Any further decrease in pressure will
damage the device!
Reduce differential pressure to
±
250 mbar to avoid damage.
Any further increase in pressure will
damage the device!
Visually inspect the flow channel
•
Contact IMT Analytics soon for
•
calibration:
www.easy-cal.com
Check measurement setup
•
Hard reset the device. To do so:
1. Disconnect any cables including
the power supply
2. Press and hold the power button
for at least 10 sec.
3. Device will shut down and restart
automatically