Operation
6
34
6.5.6
Software update
The software update file can be downloaded from the IMT Analytics website. A USB
flash drive with FAT32 formatting is required to install new software.
To update the FlowMeter software:
1. Copy the software update file to the root directory of a FAT32-formatted USB
flash drive
2. Plug the USB flash drive into the USB-A host port
3. Switch on the FlowMeter
4. Tap the Menu button to open the main menu
5. Select Settings
6. Select About
7. Select SW update
The device verifies if a valid software update file is available on the connected
USB flash drive. A progress bar is shown on the screen as follows:
If the verification was successful, the following message is shown on the screen:
If the verification was unsuccessful, an error message is shown
on the screen. See details about possible error messages
below.
8. Tap Press to start the update!
The software update starts. A progress bar is shown on the screen as follows: