Table of contents Introduction Intended use Safety instructions Representation of hazards, cautions and notes Personnel Responsibility and guarantee Service life Specifications Measurement parameters Gas standards for flow and volume measurement Power supply Battery operation Directives and approvals Device labels and symbols PC minimum requirements Start-up Individual parts in the packaging...
Page 4
Notes about changing parts 10.3 Preventive cleaning and servicing routines 10.4 Contact 11 Accessories and spare parts 11.1 Ordering address 11.2 Spare parts 12 Disposal 13 Appendix 13.1 Abbreviations and glossary 13.2 Measurement parameters and units 13.3 Index IMT Analytics AG...
Introduction Validity This documentation is valid for the product with the designation: • FlowAnalyser PF-300 PRO, REF 700.300.000 • MultiGasAnalyser OR-703, REF 500.041.000 You will find the name FlowAnalyser PRO on the rating plate on the side of the device.
• Inspiratory time, expiratory time • Inspiratory hold time, expiratory hold time • P peak • P mean • P plateau • PEEP • PF Insp (inspiratory peak flow) • PF Exp (expiratory peak flow) • Ti/T • C stat IMT Analytics AG...
Page 7
• contamination of the device • contamination of the patients lung • loss of ventilation performance • increase of dead value This product is intended to be used at elevations of up to 3000 MASL in buildings. IMT Analytics AG...
If you are aware of an incident, please contact the office listed below. Technical support Tel: +41 (0)81 750 67 10 Email: techsupport@imtanalytics.com Service life The maximum service life of the device has been specified as 10 (ten) years, provided it is handled properly in accordance with this User Manual. IMT Analytics AG...
± 5 % RH** for <10˚% RH and > 80 % RH Standard litres per minute (converted to STP conditions of 21.1°C and 1013 mbar) Tolerance related to the measured value ** Ablsoute tolerance, with steady air flow *** Non-condensing IMT Analytics AG...
Page 10
Insp. hold time Range 0 – 60 s Hold Accuracy ± 0.01 s Exp. hold time Range 0 – 90 s Hold Accuracy ± 0.01 s Tolerance related to the measured value ** Ablsoute tolerance, with steady air flow IMT Analytics AG...
Page 11
MultiGasAnalyser OR-703 (optional), trigger input (digital) for external trigger. 4.1.6 Physical data Weight: 3.2 kg Size (l × w × h): 312 × 233 × 130 mm Gas types: Air, O O, He, N , CO mixtures: Air/O , He/O IMT Analytics AG...
Page 12
Storage ambient pressure: 540 – 1100 hPa Storage and transport conditions: -10 – 60 °C (14–140 °F) 5 – 95 % RH (non-condensing) Keep dry und keep away from sunlight. IP classification: IP21 Installation categorie: Class II IMT Analytics AG...
0 °C (32 °F) according to DIN 1343 (760 mmHg) Standard condition 20/981 981 mbar 20 °C (68 °F) according to ISO 1-1975 (736 mmHg) (DIN 102) Please always use standard condition STP for ultra-low flow meas- urements. IMT Analytics AG...
The device indicates visually and audibly when the battery must be charged. Please do not store the battery in depleted state. Caution: depletion can damage the battery beyond repair! IMT Analytics AG...
USB Port to connect an USB Stick Serial interface Port to communicate with a computer Battery Conforms to CE Directives and Regulations Safety Mark for North America for PF-300 PRO Safety Mark for North America for Power Supply Manufacturer IMT Analytics AG...
Start-up Start-up Individual parts in the Picture Description packaging FlowAnalyser PRO Power supply (Order No. 700.355.000) USB cable Calibration certificate Filter Adapter set IMT Analytics AG...
(9.2. Connecting to the ventilator). Particles of dirt in the air can clog the measuring system and thus lead to incorrect measurements. The filter must be checked regularly (10.3 Preventive cleaning and servicing routines). IMT Analytics AG...
Page 19
± 1.65 % of reading or ± 0.04 L/min If work is being performed at a relatively high level of air humidity, care must be taken to ensure that no condensation forms in the device. Water can damage the sensors beyond repair! IMT Analytics AG...
Page 20
± 1.65 % of reading or ± 0.01 L/min 5.3.5 Low differential pressure The low differential pressure connectors can be used for differential pressure meas- urements with low pressures. Measuring range: -10 –10 mbar Accuracy: ± 1 % of reading or ± 0.01 mbar IMT Analytics AG...
Page 21
The high-pressure port can be used for measuring pressures from -1 to 10 bar. If a DISS-O connector is required for the port, an appropriate adapter can be ordered. Measuring range: -1 –10 bar Accuracy: ± 1 % of reading or ± 7 mbar IMT Analytics AG...
If the device is to be actuated via the RS-232 interface, various data protocols are available. FlowAnalyser PRO pin assignment (RJ-45 connector): Pin 1 +5 V (+/- 0.2 V), max 0.5 A Pins 4, 5 Pin 7 Pin 8 Pins 2, 3, 6 No connection IMT Analytics AG...
Page 23
The input is electrically isolated. Maximum 24V and 60 mA to be used on external trigger interface. 5.4.4 The USB host connector can be used to connect a memory stick for data logging and also for firmware upgrade. IMT Analytics AG...
Page 24
Start-up 5.4.5 Ethernet The Ethernet connector allows to connect the device to a LAN. The device will pro- vide an embedded web server. IMT Analytics AG...
Drag with Place index finger Moving of measurement two fingers and middle finger curves when the meas- on object and drag. urement curve is paused. IMT Analytics AG...
You can load, edit and create functions in this menu item. Applications With the menu item Applications, you are offered custom applications. If you are in need of such a solution, please contact IMT Analytics at: sales@imtanalytics.com About This menu item saves user information and device information.
Page 28
The associated settings can be found in the sub-item Trigger. You will find the settings concerning the device in the sub-menu Device. This contains various adjustment options for the screen, device settings and the various interfaces. IMT Analytics AG...
In the full-screen view you have the option of changing the parameter, the unit of measurement, the gas standard and the gas type. With the pause button you can also freeze the curve and perform measure- ments on it. IMT Analytics AG...
During installation, make sure the battery charge is at least 50 % or the device is plugged into the power supply using the power supply unit included. While installation is running, the device must not be switched off! IMT Analytics AG...
To export the test reports, a USB stick with the adapter included must be plugged into the USB host port on the device. If you require an application for your device tailored to your particular needs, please contact sales@imtanalytics.com. IMT Analytics AG...
If you press Change, a plus and a minus appear, so you can switch between the various requirements. Save applies the value selected. A gas that has not been selected properly and a gas standard that has not been selected properly can lead to measuring errors of up to 20 %. IMT Analytics AG...
Save applies the filter selected. This filtering of readings only has an impact on the values shown on the display of FlowAnalyser PRO. In the FlowLab software it is always the raw, unfiltered readings that are displayed. IMT Analytics AG...
When you have switched on the device, individual displays may devi- ate slightly from the zero point until operating temperature has been reached. Zero calibration should never be performed while the device is cold. Warming-up time is approx. 10 minutes. IMT Analytics AG...
20 and 30 L/min and press Start Oxygen Calibration. Follow the in- structions on the screen. In the second step, you must connect air with a gas flow between 20 and 30 L/min. The entire process takes about four minutes. IMT Analytics AG...
Measurements can be detrimentally affected by RF radiation, e.g., by cell phone communication. Care must be taken to ensure that MultiGasAnalyser is only operated in an EMC-specified environment. IMT Analytics AG...
A green LED indicates that the sensor is ready for operation. LED signal Continuous green light System OK Continuous blue light Anaesthetic agent exists Continuous red light Sensor error Flashing red light Please check the adapter Flashing green light Calibration of the OR sensor IMT Analytics AG...
Gas measurements should be checked regularly using a reference measuring instru- ment. For this purpose, IMT Analytics AG offers a service to recertify the measuring accu- racy of the sensor. IMT Analytics AG...
CO values by 6 %. This means that a measured mixture consisting of 5.0 vol% and 50 vol% helium is equivalent to a measured concentration of (1 0.06) * 5.0 – vol% = 4.7 vol% CO IMT Analytics AG...
For the automatic trigger to determine the limits reliably, the last 30 – 40 seconds of flow and pressure data is evaluated. Though the automatic trigger works very reliably, it may take up to 40 seconds to adapt to changed ventilation settings. IMT Analytics AG...
B: In the inspiratory channel upstream of the Y-piece FlowAnalyser PRO Ventilator Testlung Note: default trigger settings require modification. C: In the expiratory channel upstream of the Y-piece Ventilator FlowAnalyser PRO Testlung Note: default trigger settings require modification. IMT Analytics AG...
If, for example, a system has a defined leak, resulting in a continuous discharge of 3 L/min air, those 3 L/min are not included in the inspiratory volume. By entering • Base flow: on 3.0 L/min the volume calculation could be corrected in our example. IMT Analytics AG...
(> 3 L/min / < -3 L/min) can be used in this case without any problems. In such a situation it is important to ensure that the trigger is signifi- cantly above the noise of the baseline or else false triggering can be caused. IMT Analytics AG...
Page 44
Naturally, it is also important to ensure that the trigger is significantly above the noise of the baseline. Otherwise, the trigger value must be increased. In the FlowLab software it is very easy to determine with the aid of the cursor where the trigger should be placed. IMT Analytics AG...
Many ventilators do not include these small flows in the calculation of Vti. With the following trigger settings, this can also be prevented in FlowAnalyser PRO: S on this chart represents the start trigger and E represents the end trigger. IMT Analytics AG...
Page 46
1 L/min to 3 L/min, de- pending on the respiratory mode. At the same time, the pressure range during the plateau must not be bigger than 0.5 mbar. IMT Analytics AG...
Visually check the protection filter for dust or contamination. Replace the filter if con- tamination is visible. Every 12 months: Factory calibration to ensure reliable measurement. To have FlowAnalyser PRO calibrated at the manufacturer's, IMT Analytics AG, visit the website www.imtanalytics.com/easycal IMT Analytics AG...
Page 48
Remove battery Only use the spare parts recommended by the manufacturer (11 Accessories and spare parts). The battery must not be replaced with a different type. Doing so may lead to a explosion, burn or fire hazard. IMT Analytics AG...
Servicing and maintenance 10.4 Contact If you have any questions or problems please contact one of the offices listed below. 10.4.1 Manufacturer‘s name and address IMT Analytics AG Gewerbestrasse 8 CH-9470 Buchs Switzerland Tel: +41 (0)81 750 67 10 Email: sales@imtanalytics.com...
Accessories and spare parts 11 Accessories and spare parts 11.1 Ordering address IMT Analytics AG Gewerbestrasse 8 CH-9470 Buchs Switzerland Tel: +41 (0)81 750 67 10 Email: sales@imtanalytics.com 11.2 Spare parts Picture Name Article Number Battery Pack 700.348.000 Power Supply 700.355.000...
• without endangering human health • without using processes or methods that harm the environment, especially water, air, soil, animals and plants • without causing noise or odour nuisances • without having a detrimental effect on the environment or landscape. IMT Analytics AG...
Inspiratory/expiratory ratio Litre Light emitting diode Litres per second Max, max Maximum mbar Millibar (1 mbar = 10-3 bar) Minute Min, min Minimum min. At least Millimetre (1 mm = 10-3 m) Millilitre (1 mL = 10-3 L) IMT Analytics AG...
Page 53
Pplateau Plateau pressure at the end of inspiration Relative Humidity RS-232 Serial interface Ti//T Ratio of inspiratory time to time of a respiratory cycle Volt Apparent power input of the device Volts Alternating Current Volts Direct Current IMT Analytics AG...
°C, K, °F Volume Vol. HF mL, L, cf 13.2.4 Gas concentrations Measurement parameter Designation Units of measurement Gas concentration Gas concentration Partial pressure Partial pressure mbar, bar, inH O, cmH psi, Torr, inHg, mmHg, hPa, kPa, Pa IMT Analytics AG...
Page 55
O, l/cmH Inspiratory time Inspiratory hold time Ti Hold s, ms Expiratory time Expiratory hold time Te Hold Peak expiratory flow PF Exp. L/min, L/s, cfm, mL/min, mL/s Peak inspiratory flow PF Insp. IMT Analytics AG...