IMT Analytics AG
Once the software update is loaded, the device reboots automatically. This is
indicated on the screen as follows:
During the reboot, the screen turns white for a couple of seconds. After that, the
following message is displayed on the screen:
Once the update is completed, the measurement screen is displayed
9. Disconnect the USB flash drive from the FlowMeter
The device is ready for use
If the verification of a valid software update file was unsuccessful, one of the following
error messages is shown on the screen:
Error message displayed
Troubleshooting
1. Tap Retry! to try it again
2. Ensure that a USB flash drive is connected
to the USB-A host port
(5.4 Electrical
interfaces)
3. Ensure the battery charge is > 50%
4. Format the USB flash drive using the file sys-
tem FAT32, then re-download the software
update file from the IMT Analytics website,
save the software update file named Flow-
MeterUpdate.bin in the root directory of the
USB flash drive, disconnect the USB flash
drive from your computer safely (Eject) and
try to install the software again
5. Repeat the process with a different USB
flash drive
Operation
6
35