Cleaning; Trouble Shooting - IMT Analytics FlowMeter F1 User Manual

Table of Contents

Advertisement

8.4

Cleaning

8.5

Trouble shooting

IMT Analytics AG
If the screen can not be cleaned under running water, exchange it
with a new one
(9.2
Note: If the flow channel was observed to be clean in the past, contamination can
easily be recognized.
Every 12 months
Factory calibration ensures reliable measurement. To calibrate the FlowMeter by the
manufacturer, visit the website
Only use the spare parts recommended by the manufacturer
(9 Accessories and spare
For cleaning, only use agents which do not affect the functioning of the FlowMeter:
• Flow channel: dry, compressed and clean air
• Enclosure: soft, lint-free cloth, dry or moist (soap, mild cleaning agent or iso-
propyl alcohol)
For cleaning, do not use any solvent or chlorine based or abrasive detergents. Sol-
vents can permanently offset the humidity sensor.
The device is not inteded to be desinfected or reprocessed and shall therefore be
kept clean.
Using incorrect cloths can cause scratches on the device.
Fault
Possible cause
FlowMeter not
No power
turning on
Error message:
Sensor defective or extreme
Sensor Failure!
environment condition
Error code: xxxx
Error message:
Battery disconnected or de-
Battery not
fective
available!
Error message:
Measurement subsystem not
Measurement
responding
not available
Error message:
12 month since last calibration
Calibration
overdue!
Parts).
www.easy-cal.com
parts).
Trouble shooting
Connect power supply, check voltage,
check all connections, use a compli-
ant power supply as described in the
chapter specification
Follow allowed operation conditions
and restart the device. If the error
persists, contact the distributor or
manufacturer for repair of the device
Restart the device and charge the
battery for at least 2 hours. If the error
persists, contact the distributor or
manufacturer for repair of the device
Restart the device. If the error persists,
contact the distributor or manufacturer
for repair of the device
Contact IMT Analytics for calibration:
www.easy-cal.com
Servicing and care
8
41