Servicing and care
8
Servicing and care
8.1
Guidelines for servicing
and care
8.2
Notes about changing parts
8
8.3
Preventive cleaning and
servicing routines
40
Careful service in compliance with the instructions is essential for ensuring that
FlowMeter operates safely and efficiently. Only components recommended by the
manufacturer may be used.
It is absolutely essential to comply with the guidelines and service
instructions issued by the various manufacturers.
The service routines listed below may only be performed by persons
who are familiar with FlowMeter. All further repair work may only be
performed by authorised trained professionals. Please also observe
the information issued by the various manufacturers.
To ensure that your device operates with precision and reliability for as long as possi-
ble, it is essential to perform the following servicing routines regularly:
Every four weeks
For best accuracy, visually check the screens inside the flow ports. Inspect the
screens for contamination by looking through each end of the flow channel
while pointing the other end of the flow channel at a light source.
If the screen is contaminated, clean it according to the following procedure:
• Unscrew the flow port retention nut
• Remove the flow port connector
• Wash the screen under running water and dry it
• Reinsert the screen and flow port into the device
• Install and tighten the flow port retention nut
1
2
1
and the screen
2
3
3