Calibrating the Probe on InfiniiVision Oscilloscopes
Procedure
N2797A User's Guide
Allow the oscilloscope and probe to warm up for 20 minutes
before performing the probe calibration.
1
Connect the probe to one of the oscilloscope channels.
NOTE
When you connect the probe, the Calibrate Probe softkey in the Channel
Probe Menu becomes active.
2
Connect the probe to the calibration terminals. On 3000
X- series oscilloscopes, connect the probe's signal tip to the
front- panel Demo 2 / Probe Comp terminal and the ground tip
to the ground terminal.
Figure 11
Probe Calibration Terminals on 3000 X-Series Oscilloscope
3
Press the Channel on/off key to turn the channel on (if the
channel is off).
4
In the Channel Menu, press the Probe softkey.
Calibrating the Probe on InfiniiVision Oscilloscopes
Probe Calibration
2