Agilent Technologies N2797A User Manual page 45

Extreme temperature single-ended active probe
Table of Contents

Advertisement

N2797A User's Guide
j
Press broadband screen key selection.
k
Wait until the VNA beeps indicating that it has completed
the task.
l
Press the done loads screen key.
m
You have just calibrated one side of the reference plane.
13
Connect the other high quality SMA cable to the VNA's
PORT 2 as shown in
Figure 17
SMA Cable Connected to Port 2
14
Get the opposite sex of the Calibration Standards for the
next step.
15
Perform Calibration for the PORT 2 side of the Reference
plane.
a
Press the reflection screen key.
b
Connect the open end of Calibration Standard to the
available end of the PORT 2 SMA cable.
c
Select the open screen key under the Reverse group.
d
Wait until the VNA beeps indicating that it has completed
the task.
Performance Verification
Figure
17.
5
Test 2. Bandwidth
45