Agilent Technologies N2797A User Manual page 34

Extreme temperature single-ended active probe
Table of Contents

Advertisement

3
Characteristics and Specifications
Table 8
Electrical Characteristics and Specifications
Parameter
a
Bandwidth (–3 dB)
Rise Time (10% – 90%, calculated)
System Bandwidth (with Agilent oscilloscope)
Attenuation Ratio (at DC)
Input Dynamic Range
Non-Destructive Maximum Input Voltage
Offset Range
DC Offset Error (Output Zero)
Flatness (at 25° C)
Flatness Over Temperature (–40° C to +85° C)
a
Input Resistance
Input Capacitance
Probe Noise
Output Impedance
Internal Power
a Denotes warranted electrical specification at 25 °C room temperature after 20 minute warm-up. All others are typical.
34
Characteristic or Specification (where indicated)
1.5 GHz (specification)
233 ps
1 GHz (with Agilent's 1 GHz InfiniiVision or Infiniium
oscilloscope)
10:1 ± 0.5%
–8V to +8V (DC or Peak AC)
–20V to +20V
±12V
< ±1 mV
Typical 0.3 dB (10 Hz – 100 MHz)
Typical 0.8 dB (100 MHz – 1 GHz)
Typical 2.0 dB (1 GHz – 1.5 GHz)
Typical 0.3 dB (10 Hz – 100 kHz)
Typical 0.6 dB (100 kHz – 100 MHz)
Typical 0.8 dB (100 MHz – 500 MHz)
Typical 2.0 dB (500 MHz – 1 GHz)
Typical 2.5 dB (1 GHz – 1.5 GHz)
1 MΩ ± 3% (specification)
1 pF
< 2.5 mV
(referred to input)
rms
50Ω
Agilent AutoProbe interface from oscilloscope
(InfiniiVision and Infiniium)
N2797A User's Guide