N2797A User's Guide
Figure 15
Probe Connected to Probe Adapter (not to scale)
Calibrating a Reference Plane
To get a reliable measurement from the VNA you must
calibrate a reference plane so that the VNA knows where the
probe under test is located along the transmission line.
7
On the VNA, press the [Cal] key.
8
Press the cal menu screen key.
9
Press the full 2 port screen key.
10
Connect the PV fixture to the Port 1 on the VNA using a high
quality SMA cables. Connect the cable to the pincher side of
PV fixture as shown in
11
The calibration reference plane is at the other end of PV
fixture.
Performance Verification
Test 2. Bandwidth
Figure 16
on page 44.
5
43