Operation
Model 8902A
PROCEDURES
Entering Special Functions.
To
use a special function, key in the corresponding code, then select SPCL.
Special Display.
To display the user-requested modes of special functions prefixed
1 through 10, select SPCL alone
one time. The digit position (noted beneath the display) corresponds to the special function prefix,
and the number displayed in that position corresponds to the special function suffix.
Special Special Display.
To determine the actual instrument settings corresponding to special functions prefixed
1
through
10, press the SPCL key alone once while the Special Display
is active. (If the Special Display is not
active, press the SPCL key twice to get this display.) The digit position corresponds to the special
function prefix, and the number displayed in that digit corresponds to the special function suffix.
HP-IB PROGRAM CODES
All special fiinction codes are formed by the special function number followed by the code "SP".
For example, the code for Special Function 7.3 is "7.3SP".
For special function codes with zero as a suffix, force a single space if the zero is to be ommitted.
For example, "1O.OSP" could be
"IO.
SP"
EXAMPLES
Entering Special Functions.
To display the frequency of the signal in the IF (Special Function 34):
LOCAL
(keystrokes)
@@
(program codes)
34.0
SP
Code
Function