HP 8902A Quick Reference Manual page 218

Measuring receiver
Hide thumbs Also See for 8902A:

Advertisement

Operation
M o d e l
8902A
Clear Message Check
This
check determines whether
or
not the Measuring Receiver properly responds to the Clear
message.
This
check assumes that the Measuring Receiver
i s able to handshake, recognize i t s
own
address, make the remote/local changes. and receive Data messages. Before beginning
this
check, select
40.0 SPCL.
Description
Send the first part of the Remote
message (enabling the Measuring
Receiver
to
remote).
Address the Measuring Receiver to listen
(completing the Remote message). Then
send a Data message that sets the
Measuring Receiver's tuning to manual.
Check that the Measuring Receiver's
REMOTE and LISTEN annunciators are
on and that the AUTO TUNING light is
off.
Send the Clear message (setting the
Measuring Receiver's tune mode back to
automatic).
Check that the Measuring Receiver's
REMOTE and LISTEN annunciators are
on and that the AUTO TUNING light is
on.
HPL
rem
7
wrt 714,"MZ"
clr 714
BASIC
REMOTE 7
OUTPUT 714;"MZ"
RESET 714
CLEAR 714