Operation
Model
8902A
SPCL
37.0
-
-
37.1
37.2
37.3
37.4
-
37.5
37.6
37.9
Summary
of
Special Function
37
Sbecial Function Describtion
~~
Selects the Automatic Cal Factor mode. This
en-
ables the Measuring Receiver to automatically se-
lect
cal factors based on those entered using Special
Function 37.3.
Selects the Manual Cal Factor Mode. This enables
the user to override the automatic selection of cal
factors.
Procedure :
To enter a manual cal factor, select:
37.1 SPCL, cal factor value,
O h
CAL FACTOR (Blue
Key, MHz)
Displays the status
of the cal factor mode
(0 =Automatic,
1 =
Manual).
Prepares
to enter a reference cal factor or a fre-
quencyhl-factor pair into the appropriate table in
memory.
Procedures:
To enter the reference cal factor, select: 37.3 SPCL,
cat factor value,
O h
CAL FACTOR (Blue Key, MHz).
To enter each frequencykal-factor pair, select:
37.3 SPCL, frequency value (in MHz), MHz key, cal
factor value,
O / O
CAL FACTOR (Blue Key, MHz). (It's
unnecessary to wait until the values are displayed be-
fore continuing with subsequent entries.)
(See examDles in this section.)
Displays the number of entries in the table.
Prepares to recall the reference cal factor from the
table.
Procedure:
To display the reference cat factor, select:
37.5 SPCL,
O/O
CAL FACTOR (Blue Key, MHz).
Prepares to recall the next frequencykal-factor pair
from the table.
Procedure
:
To display the frequency, select:
37.6 SPCL, DISPLAY FREQ (Blue Key, kHz).
To display the cal factor, select:
O/O
CAL FACTOR (Blue Key, MHz).
37.6 SPCL sequentially accesses frequencykal-factor
pairs.
(See example in this section.)
Clears the table of cal factors from memory.
To
enter
a manual cal factor, use 37.1 SPCL. To enter a new
table
of
calibration factors, use
37.3 SPCL.
Comments
0
This mode
is
selected upon instrument power-up and
with Automatic Operation.
0
Can be keyed from any Measurement mode.
0
If
no cal factor is entered, the Measuring Receiver uses
the cal factor recalled last from the table as the manual
entry.
0
The manual cal factor remains in effect when the
Measuring Receiver is retuned.
0
The automatic
cat
factor table
is
not affected
by
any
manual cal factor entries.
0
Can be keyed only after selecting RF POWER.
0
Can
be
keyed from any Measurement mode.
0
If
an error is made during entry of any value, select
CLEAR and immediately re-enter the correct value.
0
To replace an existing cal factor entry, enter the new
cal factor at the frequency previously used in the entry.
The Measuring Receiver will automatically save the entry
in the correct order in the table.
0
To remove an existing
cal
factor from the table in
memory, enter 0% as the cal factor at the frequency
previously used in the entry.
0
Can be keyed from any Measurement mode.
0
The number of table entries displayed also includes
the reference calibration factor.
0
Can
be
keved from anv Measurement mode.
0
This special function configures the Measuring Re-
ceiver to use manual cal factors with the reference cal
factor from the table as the manual entry. To re-enter the
Automatic Cal Factor mode, select AUTOMATIC OPER-
ATION or 37.0 SPCL.
0
Can be keved only after selectina RF POWER.
0
The table in memory is accessed sequentially (the low-
est frequency accessed first). To access the lowest fre-
quency in the table, select 37.5 SPCL before selecting
37.6 SPCL.
0
This special function configures the Measuring Re-
ceiver to use manual cal factors with the cal factor re-
called last from the table as the manual entry. To re-
enter the Automatic Cal Factor mode, select AUTO-
MATIC OPERATION, then reselect RF POWER.
An attempt to read beyond the end of the table results
in Error 26.
0
Displays Error
15
if
automatic cal factors were being
used.
Can be keved from anv Measurement mode.
3-78