Download Print this page

HP 8902A Quick Reference Manual page 205

Measuring receiver
Hide thumbs Also See for 8902A:

Advertisement

Model
8902A
Opera
tion
AM and FM Calibration Check
16. Connect
the input of the sensor module to the AM/FM CALIBRATION
OUTPUT
connector
(See Figure 3-8.)
HP
8902A
MEASURING
RECEIVER
I,
0
AMIFM
CALIBRATION
OUTPUT
Figure 3-8. A M and FM Calihralion Chc.c.k StVirp
17. Press the AM key then the CALIBRATE key. After approximately 20 seconds, the AM
calibration factor should be displayed, The displayed value should be between
99.00
and
10 1
.ooo/o.
18. Press the
FM key then CALIBRATE
key. After approximately 20 seconds, the
FM calibration
factor should
be displayed. The displayed value should be between 99.00 and 101.OOo/o. Press
the CALIBRATE key again to turn o f f FM calibration. (If you remove the sensor with the
calibrator still on it might return Error 08.)