HP 8902A Quick Reference Manual page 385

Measuring receiver
Hide thumbs Also See for 8902A:

Advertisement

Model 8902A
Operation
COMMENTS
Default Condition.
When the instrument is first powered up o r when Automatic Operation o r Instrument Preset is
selected, R F and IF Filters are placed in the automatic selection mode.
Automatic Retuning.
When a new IF is selected, the instrument immediately retunes.
Increased Sensitivity and Selectivity.
Many measurements can typically be made on very low-level signals. Using the narrow bandwidth o f
the 455 kHz IF for signal frequencies less than 2.5 MHz (but in the manual down-convert mode) o r
greater than 10 MHz increases the sensitivity and selectivity o f the Measuring Receiver. (There are,
however, restrictions o n modulation, and measurement accuracy may decrease.) Manual deletion o f
RF input attenuation o r insertion of gain also increases sensitivity. (Refer to
R1:
Input A ((cwuation
and Gain o n page 3-265.)
3-26
1