Removing and Replacing the Acquisition Board

1
2
3
S-Series Oscilloscopes Service Guide
Remove the air duct from the acquisition board by lifting it straight up.
Reposition the oscilloscope so you can access the front panel, and remove the
four M2 screws next to each BNC. When replacing, torque to 3 in-lbs using a
power driver and T-6 bit.
Remove the two M3 screws that connect the heat sink to the chassis. Do not
remove the other two screws.
5
Replacing Assemblies