3 Testing Performance
Full performance verification for the S-Series oscilloscopes consists of three main
procedures:
Performing the internal oscilloscope self-tests to ensure that the measurement
1
system is functioning properly. To perform the self-tests, click Utilities > Self
Test.... Then select Scope SelfTest from the Available Self Test drop-down list
box, click Start, and follow the on-screen instructions. If any of the self tests fail,
ensure that the failure is diagnosed and repaired before calibrating and testing
performance.
Calibrating the oscilloscope, as described in
2
Testing the oscilloscope to ensure that it is performing to specification. This
3
chapter describes these performance test procedures.