4
Troubleshooting

Checking probe power outputs

Probe power outputs are on the front panel, surrounding each BNC input.
Refer to the following figure to check the power output at the connectors. Measure
the +5 V, -5 V, +12 V, and -12 V voltages with respect to the Probe Comp ground
terminal on the front panel.
Do not try to measure voltages at pins 3 through 7.
Any failure is likely caused by a problem with the front panel board.
70
Pin
Supply
1
+5 V
2
–5 V
3
Offset
4
Data
5 & ring
Probe ID
6
Clk
7
Unused
8
–12 V
9
+12 V