Setting Up The Bios - Keysight Infiniium S Series Service Manual

Hide thumbs Also See for Infiniium S Series:
Table of Contents

Advertisement

4
Troubleshooting

Setting up the BIOS

If the BIOS settings become corrupt, the Infiniium oscilloscope PC motherboard
will not recognize the hard drive and the oscilloscope may not start. To configure
the motherboard BIOS parameters to the default settings, follow these steps:
1
2
3
Press DEL to enter SETUP, F12 to select boot device.
4
5
66
Connect the power cable to the oscilloscope.
Connect the external keyboard to the rear panel.
Press [Del] as soon as the following prompt appears on the bottom of the
screen:
Go to F3 Optimized Defaults and press [Enter]. Select Yes to load the defaults, then
press [Enter].
Go to the Save and Exit Setup option and press [Enter] to save and exit the setup.
Select Yes to save and exit, then press [Enter].
S-Series Oscilloscopes Service Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents