Endo; Ov-V - EDAN DUS 60 User Manual

Hide thumbs Also See for DUS 60:
Table of Contents

Advertisement

DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual
R. OV-L
Right Ovary Length
R. OV-W
Right Ovary Width
R. OV-H
Right Ovary Height
L. FO-L
Left Follicle Length
L. FO-W
Left Follicle Width
R. FO-L
Right Follicle Length
R. FO-W
Right Follicle Width
CX-L
Cervix Length
The ratio of Uterus Length
UT-L/CX-L
and Cervix Length
Table 9-1 Gynecology Measurement and Calculation Items in B Mode
9.1.1. UT
To measure UT:
1. In the gynecology measurement menu, roll the trackball to highlight UT and press Set.
2. Take three measurements, UT-L, UT-W and UT-H, in the method of distance measurement.
Reference Section 6.6.1, Generic Measurements in B Mode
3. After the three measurements are taken, the result of UT will be displayed in the measurement
result window. You can measure a maximum of one group of data.

9.1.2. Endo

To measure endometrium:
1. In the gynecology measurement menu, roll the trackball to highlight Endo and press Set.
2. Measure Endo, in the method of distance measurement.
Reference Section 6.6.1, Generic Measurements in B Mode
3. The result of Endo will be displayed in the measurement result window. You can measure a
maximum of one group of data.

9.1.3. OV-V

The measurement of OV-Vol includes L.OV-Vol and R.OV-Vol.
To measure L.OV-Vol:
1. In the gynecology measurement menu, roll the trackball to highlight OV-Vol, and then
highlight the secondary menu L.OV-Vol, press Set.
To Purchase, Visit
Avobus.com
Gynecology Measurements and Calculations
Distance (mm)
UT-L/CX-L Ratio
- 118
-
or call
1-800-674-3655
DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual
2. Take three measurements, L.OV-L, L.OV-W and L.OV-H, in the method of distance
measurement.
Reference Section 6.6.1, Generic Measurements in B Mode
3. After the three measurements are taken, the result of L.OV-Vol will be displayed in the
measurement result window.
To measuree R.OV-Vol:
1. In the gynecology measurement menu, roll the trackball to highlight OV-Vol, and then
highlight the secondary menu R.OV-Vol, press Set.
2. Take three measurements, R.OV-L, R.OV-W and R.OV-H, in the method of distance
measurement.
Reference Section 6.6.1, Generic Measurements in B Mode
3. After the three measurements are taken, the result of R.OV-Vol will be displayed in the
measurement result window.
9.1.4. FO
The measurement of FO includes L. FO and R. FO.
To measure L. FO:
1. In the gynecology measurement menu, roll the trackball to highlight FO, and then highlight
the secondary menu L. FO, press Set.
2. Take two measurements, L. FO-L and L. FO-W, in the method of distance measurement.
Reference Section 6.6.1, Generic Measurements in B Mode
3. After the two measurements are taken, the result of L. FO will be displayed in the
measurement result window.
To measure R. FO:
1. In the gynecology measurement menu, roll the trackball to highlight FO, and then highlight
the secondary menu R. FO, press Set.
2. Take two measurements, R. FO-L and R. FO-W, in the method of distance measurement.
Reference Section 6.6.1, Generic Measurements in B Mode
3. After the two measurements are taken, the result of R. FO will be displayed in the
measurement result window.
Gynecology Measurements and Calculations
- 119
-

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents