EDAN DUS 60 User Manual page 68

Hide thumbs Also See for DUS 60:
Table of Contents

Advertisement

DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual
Figure 10-2 Small Parts Worksheet
The diagnosis editing column displays the cursor "І", and you can enter diagnosis information.
To print the report:
Press Print in the Small Parts Worksheet.
Printing reference Section 5.8, Printing.
10.3. Others
Select Others to enter another application measurement.
Small Parts Measurements and Calculations
- 126
-
DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual
Chapter 11 Urology Measurements and Calculations
11.1. Measurement and Calculation
The urology examination is usually in the B mode.
1. Press Exam and select Urology, and then press Set.
2. Press
to enter the B mode.
3. Press Measure to activate the measurement function. The system displays the measurement
menu shown as below.
Figure 11-1 Urology Measurement and Calculation Menu
Label
Description
Method
Residual Urine Volume
RUV (mL) = 0.7x RUV-L (mm) × RUV-W (mm) × RUV-H (mm)
RUV
(mL or L)
/1000
RUV-L
Residual Urine Length
RUV-W
Residual Urine Width
Distance (mm)
RUV-H
Residual Urine Height
Prostate Volume
PV
PV (mm
3
3
3
(mm
, cm
, or dm
)
PV-L
Prostate Length
PV-W
Prostate Width
Distance (mm)
PV-H
Prostate Height
Predicted
Prostate
PPSA
PPSA (ng/mL) = 0.12 × PV
Specific Antigen Density
Serum
of
Prostate
SPSA
Key in SPSA (ng)
Specific Antigen
Prostate
Specific
PSAD
PSAD (ng/mL) = SPSA (ng)/ PV (mL), (0.01ng ≤SPSA≤100ng)
Antigen Density
Table 11-1 Urology Measurement and Calculation Items
To determine the residual urine volume or prostate volume, take three measurements: length,
width, and height. The system calculates the volume.
Urology Measurements and Calculations
3
) = 0.52 x PV-L (mm) × PV-W (mm) × PV-H (mm) /1000