EDAN DUS 60 User Manual page 61

Hide thumbs Also See for DUS 60:
Table of Contents

Advertisement

DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual
LVLs
Left Ventricular Length
HR
Heart Rate
EDV
End Diastolic Volume
ESV
End Systolic Volume
SV
Stroke volume
CO
Cardiac Output
EF
Ejection fraction (B mode)
SI
Stroke Index
CI
Cardiac Index
BSA
Body Surface Area
Table 8-8 Measurement and Calculation Items with Bullet Formula
4. Modified SIMPSON formula:
NOTE: d: end diastolic; s: end systolic
Label
Description
Left Ventricular Fractional Area of
LVAMd
Mitral Valve
LVLd
Left Ventricular Length
Left Ventricular Fractional Area of
LVAPd
Papillary Muscles
LVAMs
Left Ventricular Anterior Wall
LVLs
Left Ventricular Length
Left Ventricular Fractional Area of
LVAPs
Papillary Muscles
HR
Heart Rate
EDV
End Diastolic Volume
ESV
End Systolic Volume
SV
Stroke volume
CO
Cardiac Output
EF
Ejection fraction (B mode)
To Purchase, Visit
Avobus.com
Cardiology Measurements and Calculations
Distance (mm)
Key in (bpm)
EDV (mL)= (5/6)×LVLd (mm)×LVAMd (mm
/1000
ESV (mL)= (5/6)×LVLs (mm)× LVAMs (mm
/1000
SV (mL)=EDV (mL)-ESV (mL)
CO (L/min)= SV (mL) × HR (bpm)/1000
EF (No unit)= SV (mL)/ EDV (mL) x 100%
2
SI (No unit)= SV (mL)/ BSA (m
)
2
CI (No unit)= CO (L/min)/ BSA (m
)
2
Calculate by the selected formula (m
)
Method
2
2
2
Ellipse Area (mm
, cm
, or dm
)
Distance (mm)
2
2
2
Ellipse Area (mm
, cm
, or dm
)
Distance (mm)
2
2
2
Ellipse Area (mm
, cm
, or dm
)
Key in (bpm)
*1
SV (mL)=EDV (mL)-ESV (mL)
CO (L/min)= SV (mL) x HR (bpm)/1000
EF (No unit)= SV (mL)/ EDV (mL) x 100%
- 112
-
or call
1-800-674-3655
DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual
SI
Stroke Index
CI
Cardiac Index
2
)
BSA
Body Surface Area
Table 8-9 Measurement and Calculation Items with Modified SIMPSON Formula
2
)
*1
=
(
)
EDV
mL
LVLd
=
(
)
(
ESV
mL
LVLs
5. Other measurement and calculation items:
Label
Description
LVET
Left Ventricular Ejection Time
FS
Fractional Shortening
Mean Velocity Circumferential Fiber
MVCF
Shortening
8.2.1. LV
LV measurement is as below.
Single plane ellipse (S-P Ellipse)
Measurement items:
LVLs, LVALs, LVLd, and LVALd
To measure LV:
1. In the B-cardiac measurement menu, roll the trackball to highlight LV. Then select S-P
Ellipse and press Set.
2. During end systolic, measure LVLs and LVALs respectively. The system calculates and
displays the result of ESV.
3. During end diastolic, measure LVLd and LVALd respectively, in the method of generic B
mode distance measurement and generic B mode ellipse area measurement respectively. The
system calculates and displays the results of EDV, SV, and EF.
Dual plane ellipse (B-P Ellipse), Bullet, and Modified Simpson (Mod. Simpson)
The operations in these methods are similar to those in the single plane ellipse method. Please
refer to the corresponding B mode generic measure method for details, and refer to the prompt
instruction to help you.
Cardiology Measurements and Calculations
SI (No unit)= SV (mL)/ BSA (m
CI (No unit)= CO (L/min)/ BSA (m
Calculate by to the selected formula (m
{
×
×
+
×
+
(
)
9 /
4
(
2
)
2
(
2
)
mm
LVAMd
mm
LVAPd
mm
{
×
×
2
+
×
2
+
)
9 /
4
(
)
2
(
)
mm
LVAMs
mm
LVAPs
mm
Method
Time (ms)
FS (No unit)={ LVIDd (mm)- LVIDs (mm)}/
LVIDd (mm) x 100%
MVCF (No unit)= { LVIDd (mm)- LVIDs
(mm)}/ {LVIDd (mm) x ET (ms)/1000}
Table 8-10 Other Measurement and Calculation Items
- 113
-
2
)
2
)
2
)
}
×
(
2
)
(
2
)
/
1000
LVAMd
mm
LVAPd
mm
}
2
×
2
(
)
(
)
/
1000
LVAMs
mm
LVAPs
mm