Acquisition Started but Viewer Contents Do Not Update

In LightField, live data is normally displayed in the Experiment workspace viewer as it is
being acquired in either Preview or Acquire mode. If the viewer is not being updated
and acquisition is occurring, determine if there is a filename in the top row of the viewer
For example, in
If a filename is displayed, click on the camera icon. The existing file data will be cleared
from the viewer and live data will then be displayed.
Figure E-1:
E.2

Baseline Signal Suddenly Changes

A change in the baseline signal is normal if the temperature, gain, or speed setting has
been changed.
If this occurs when none of these settings have been changed, there may be excessive
humidity in the camera vacuum enclosure.
Turn off the camera and contact Teledyne Princeton Instruments Customer Support.
Refer to
PIXIS System Manual
Figure
E-1, the filename 2010 May 11 13_13_49.raw is displayed.
Typical Acquisition Display
Contact Information
on page 154 for complete information.